Cassini TIMs can be diagnosed with the Saturn application. Follow the steps below to perform a full system Diagnostics. Diagnostics can be run at any time.



Requirements:
  • Cassini Diag Kit with Diag/Cal Interface Plate, Adapters and Cables (configuration specific)

To Diagnose Cassini with Saturn:
  1. Launch Saturn and wait for "Tester Loaded." shown in the message window. Start a 20 minute timer to allow instruments to warm up.
  2. Unlatch the Fixture if it is attached and Latch the Diag/Cal Interface plate from the Cassini Diag Kit.
  3. From Saturn app, choose apps > browse tester and choose check button and then confirm the Fixture is similar to "<Company Name> Bottomplate ##" is active. Replace "<Company Name>" with the name of the company.
  4. From Saturn app, choose apps > cassini, select cal exec and choose Launch App from the right mouse button menu.
  5. Choose File > Diag List and select the appropriate Diag list (likely "<Company Name> Rev A").
  6. Select ALL service plans (CTL+A or CTL+left mouse button click) and choose choose Run Selected from the right mouse button menu.
  7. If any FAIL results are found, follow the steps below to Send Diagnose Results to support@roos.com. If all results PASS, proceed To Calibrate Cassini 16 with Saturn below.
  8. Close the cal exec window.

To View the Worksheet from Cal Exec and Save Csv:
  1. Open the Cal Exec results worksheet, click on Actions > Results, to open the worksheet window.
  2. (Optional) Choose Data > Save Csv to convert the data in the worksheet to a .csv file in the ~\RiApps\data directory. RITdb.datalog is always saved to Guru.

To Send Diagnose Results to Roos Instruments' Support:
  1. After collecting Diag results from the Cal exec, choose Apps > Guru Browser from Guru.
  2. Choose RITdb.datalog from the Key1: ri.sys.ObjClass dropdown menu.
  3. Under Key 2, choose ri.sys.Type and select Validate.
  4. Select all the recent results in the object table and choose Export > As Guru File from the right mouse button menu.
  5. Choose OK to close the Comments window, choose a location and enter a file name and choose Save.
  6. Attach the file to an email to support@roos.com and include the Tester's serial number, reason that Diags were run, and a brief description of how to recreate the issue.


Figure 1: Example PMeter Connections
Image.jpegImage.jpeg


Figure 2: Power Sensor N to SMA Adapters (Male - Female)



PrintEmail Link
https://roos.com/docs/RBEH-DJ2QPX